Beurer IPL 7500 Mode D'emploi page 63

Réduction permanente de la pilosité
Masquer les pouces Voir aussi pour IPL 7500:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Table 2
For all me equipment and me systems
Guidance and manufacturer's declaration – electromagnetic immunity
Guidance and manufacturer's declaration – electromagnetic immunity
IPL 7500 is intended for use in the electromagnetic environment specified below. The customer or the user of IPL 7500 should assure that it is
used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interrup-
tions and voltage variations on
power supply input lines IEC
61000-4-11
Power frequency (50/ 60 Hz)
magnetic field IEC 61000-4-8
NOTICE: U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV input/output lines
± 1 kV differential mode
± 2 kV common mode
< 5 % U
(> 95 % dip in U
T
T
for 0,5 cycle
40 % U
(60 % dip in U
T
T
for 5 cycles
70 % U
(30 % dip in U
T
T
for 25 cycles
< 5 % U
(> 95 % dip in U
T
T
for 5 sec
3 A/m
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV input/output lines
± 1 kV differential mode
± 2 kV common mode
)
< 5 % U
(> 95 % dip in U
T
for 0,5 cycle
)
40 % U
(60 % dip in U
T
for 5 cycles
)
70 % U
(30 % dip in U
T
for 25 cycles
)
< 5 % U
(> 95 % dip in U
)
T
T
for 5 sec
3 A/m
63
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic materi-
al, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
)
Mains power quality should be that of a typical
T
commercial or hospital environment. If the user
)
of IPL 7500 requires continued operation during
T
power mains interruptions, it is recommended
)
that IPL 7500 is be powered from an uninter-
T
ruptible power supply or a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.

Publicité

Table des Matières
loading

Table des Matières