Power frequency
3 A/m
(50Hz) magnetic
field IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Guidance and manufacturer's declaration - electromagnetic immunity
The MB10 is intended for use in the electromagnetic environment specified below. The customer or
the user of the MB10 should assure that it is used in such an environment.
IMMUNITY test
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
100029_MEDEL CONNECT CARDIO MB10_USER MANUAL_REV.00_20.06.2016.indd 189
IEC 60601 test level
Not applicable
Not applicable
Compliance level
Electromagnetic environment -
Portable and mobile RF commu-
nications equipment should be
used no closer to any part of the
MB10 including cables, than the
recommended separation dis-
tance calculated from the equa-
3 Vrms
tion applicable to the frequency
of the transmitter.
Recommended separation dis-
tance
3 V/m
d = 1,2 P
d = 1.2 P 80 MHz to 800 MHz
d = 2.3 P 800 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the trans-
mitter manufacturer and d is the
recommended separation Dis-
tance in metres (m).