The following settings do not change when you recall default settings.
Language Option
Saved Settings
Saved waveform
Saved Reference Waveforms
Calibration Data
Horizontal System
Press Time button to enter horizontal system menu, use the direction keys to change the
horizontal scale (time base) and the horizontal trigger position. When you change the
horizontal scale, the waveform will expand or contract to the screen center.
1. SEC/DIV Knob: Used to change the horizontal time scale so as to magnify or
compress the waveform horizontally. If the waveform acquisition is stopped (by using
button), press Time button and Up or Down buttons to expand or compress the
waveform.
2. Horizontal Position Knob: Used to control the trigger position against the screen
center. Press Time button and Right or Left buttons to move the waveform right or left.
The key resolution vary according to time base. Press "AUTO" key can make the
horizontal position return to zero.
3. Mode: Y-T, X-Y, Roll, Scan.
Y-T: YT format shows the vertical voltage in relation to time (horizontal scale). Press
Time->Mode to set.
(amplitude)
Vertical (time)
Format
Position
SEC/DIV
On or Off
Source
Slope
Mode
Level
Bandwidth Limit
Coupling
Probe Attenuation
Position
VOLTS/DIV
±4div
YT
0.00s
500μs
Off
CH1
Rising
Auto
0.00v
Unlimited
AC
1X
0.00div (0.00V)
1V