should be inserted into the Gen Out output terminal, and the probe grounding clamp
should be clamped on the metal outer ring of the Gen Out output terminal.
Recommended input ~2V@1KHz peak-peak square wave.
3) Press the [Auto] button and you should see within a few seconds a square wave
of about 2V peak-to-peak at 1KHz in the display. Repeat the steps to observe CH2.

Probe Check

Safety
When using the probe, keep your fingers behind the guard on the probe body to avoid
electric shock. Do not touch metallic portions of the probe head while it is connected to a
voltage source. Connect the probe to the oscilloscope and connect the ground terminal to
ground before you start any measurements.
Manual Probe Compensation
Upon the first connection of a probe and an input channel, you should manually perform
this adjustment to match the probe to the input channel. Uncompensated or
miscompensated probes may lead to errors or faults in measurement. To adjust the
probe compensation, follow the steps below.
1. Press Channel button to enter channel setting menu, Set the Probe option
attenuation in the channel menu to 10X. Set the switch on the probe to 10X and
connect the probe to Channel 1 on the oscilloscope. If you use the probe hook-tip,
the hook end should be removed, the probe pin should be inserted into the Gen Out
output terminal, and the probe grounding clamp should be clamped on the metal
outer ring of the Gen Out output terminal. The oscilloscope with signal generator
function needs to set output signal as 2V @ 1KHz square wave; The Gen Out
terminal of the oscilloscope without signal generator function automatically outputs
2V @ 1KHz square wave. Press the [Auto] button.
2. Check the shape of the displayed waveform.
Compensated correctly
Overcompensated
Undercompensated
55