Flip the Back Shell

Warnings:
- DO NOT perform any tasks without first reading and understanding all of the safety information and provided instructions.
- DO NOT remove or install the AGILITY while the individual is seated in the wheelchair.
- DROPPING, PINCHING, CRUSHING RISK: The AGILITY with Direct Mount Hardware is heavy. To avoid injury, hold securely and use
extreme caution when installing, adjusting, removing or otherwise handling the AGILITY with Direct Mount Hardware. Ask for
assistance if needed.
Remove the AGILITY Only (Not the Direct Mount Hardware)
If the Direct Mount Hardware and AGILITY are installed on the wheelchair, follow the instructions below to remove, flip and
reinstall the AGILITY without removing the Direct Mount Hardware from the wheelchair.
1. Remove the four nylon-insert lock nuts and washers.
2. Remove the AGILITY from the wheelchair, but do not remove the Direct Mount
Hardware from the wheelchair.
3. Replace the four washers and nylon-insert lock nuts on the bolts.
Reposition the AGILITY Cover and Accessories
4. Remove the AGILITY Cover from the back shell.
Note: To remove and install the cover and to adjust the air insert(s), refer to the
ROHO AGILITY Back System Operation Manual.
5. Remove any attached headrest and/or UniTrack from one end of the AGILITY back
shell and then reattach to the other end of the AGILITY back shell.
UniTrack or headrest: Refer to the instructions provided with the product.
6. Reinstall the AGILITY Cover on the back shell. Open the top flap of the AGILITY
Cover so you will have access to the bolts.
Install AGILITY in the Flipped Position
7. Remove the lock nuts and washers from AGILITY.
8. With the AGILITY back shell flipped so the other end is in the top position, install
AGILITY on the Direct Mount Hardware that is still attached to the wheelchair:
a. Insert the four bolts into the slots in the Direct Mount Hardware. Then attach
the four washers and lock nuts on the bolts, and tighten them just enough to
allow the AGILITY to slide up and down.
b. Position the AGILITY as prescribed. Refer to the height adjustment
instructions in this manual.
c. Close the top flap on the AGILITY Cover. Make sure all inflation tubes are
correctly tucked in the cover's side pockets.