Philips PSD1211 Mode D'emploi page 16

Masquer les pouces Voir aussi pour PSD1211:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 59
ENGLISH
16
Light and sound signals during use
Overheating signal
Overheating of the device.
1
If the device measures a too high device
temperature, the Philips BlueControl switches
off for safety reasons.
, The LED indicator lights up solid orange and
a long 3-second tone indicates that the Philips
BlueControl switches off.
2
If you try to restart the Philips BlueControl
before it has cooled down, the LED indicator
flashes orange for 5 seconds. Cooling down
may take several minutes.
Overheating of the skin.
1
If the device measures a too high skin
temperature, the Philips BlueControl
temporarily switches off for safety reasons.
, The LED indicator flashes blue slowly until the
skin temperature has cooled down, the Philips
BlueControl switches on again automatically.
Battery level signals
When you press the on/off button and the LED
indicator lights up solid red and you hear 3 beeps,
the battery of the Philips BlueControl is empty.
1
You have to charge the battery before you
can start a new treatment. Charge for at
least 45 minutes to be able to complete one
full treatment.

Publicité

Table des Matières
loading

Table des Matières