Preparing the calibration check
It is possible for you to check the calibration of the laser. To do this, position
the device midway between 2 walls, which must be at least 5 metres apart.
Switch the device on. The best calibration results are achieved if the device
is mounted on a tripod. IMPORTANT: The automatic sensor must be active
(auto/man. LED is off).
1. Mark point A1 on the wall.
2. Turn the device through 180° and mark point A2. You now have a
horizontal reference between points A1 and A2.
1.
A1
2.
A1
Performing the calibration check
3. Position the device as near as possible to the wall at the height of point
A1. Now adjust the device in the X axis.
4. Turn the device through 180° and mark point A3. The difference between
points A2 and A3 is the tolerance for the X axis.
5. To check the Y and Z axis, repeat steps 3 and 4.
3.
A1
4.
If points A2 and A3 are more than 1 mm / 10 m apart on either the X
!
or Y axis, the device is in need of adjustment. Contact your authorised
dealer or else the UMAREX-LASERLINER Service Department.