Fault reporting:
Note
If a fault is reported, the electronics will give off bleep tones to alert you. If one of the above mentioned faults appears on your
display, you must not use the machine and remove the plug from the socket. Contact your dealer for more guidance.
Reading out the Counters:
+
A
B
+
C
D
Note
This timer cannot be reset to "0" .
Error!
Unknown control panel .
Error!
Internal timer is defect .
Error!
Communication fault switch part .
In order to read out the counter of the tubes, the "stop" button and the "body cooling on" button must be pressed
simultaneously for three seconds .
Timing indicator x 10 = working hours tubes
Example:
A . 759 hours is indicated as 75 .
B . 9760 is displayed alternately with 1 second intervals as: 9 . - 76 . - 9 . - 76 .
By pressing the "body cooling off" button for 10 seconds, the counter is reset at 0 hours .
The counter for the entire machine gives the total number of hours, including the ventilation time . By pressing the
"stop and body cooling off" buttons simultaneously for three seconds, the timer can be read off the display .
Timing indicator x 10 = working hours tubes
Example:
C . 759 hours is indicated as 75
D . 9760 is displayed alternately with 1 second intervals as: 9 - 76 - 9 - 76