6.3 Air Filter Maintenance

For an optimum operation of the device, the air filter can be cleaned by the user at least once
every two weeks or more often if this device is operated in a dusty environment. It is recommended
for the filter to be replaced with a new one after 6 months of use of the device; however, depending
on the air environment, the replacement time may vary.
CAUTION
:
NOTE: Please follow national requirements to dispose the unit properly.
Do not wash or dry the mask or air tubing at a temperature above 70°C
(160°F)
Do not use any cleaner containing fragrance or conditioners as they will
leave a residue.
The mask must not be re-used by another person. This is to avoid the risk
of cross-infection.
Dirty air filter may cause high operating temperatures that affect device
performance. Ensure the air
filter is cleaned and fitted at all times.
1. Remove the dirty filter from the enclosure on the rear of the
device.
2. Wash the filter in warm water with a mild detergent, and rinse
with water. Allow the filter to air dry completely before
reinstalling. DO NOT use a filter that is not completely dry. If the
filter is torn, replace it.
3. Reinstall the filter.