English
(8) Check Value (c
Press "MENU" button to select [c XXX] menu, which displays a check value. Your home
English
care provider may ask for this information.
(9) Turn off the Device
Remove the power cord from the electrical outlet, and disconnect power cord from the
power socket on the back of device.
NOTE: Once the setting is confirmed, press "MENU" button. Otherwise, the device will
4.4 Flowchart of Menu settings
Enter the user's menu mode by pressing the "MENU" button.
In each setting, when the preferred value has been selected, press "MENU" for confirmation and
press "MENU" again to enter next selection.
NOTE: For physicians, please refer to a separated "Physician's Additional Instruction"
XXX)
automatically go back to standby without saving the modification if no action is
taken in 20 seconds.
Ramp
(Ramp Time)
Ramp P
(Ramp Starting Pressure)
P
(Therapy Pressure)
Alt.
(Altitude Compensation)
TM
(Total Meter)
CM
(Compliance Meter)
CM 4 hr
(Session Counter)
c
(Check Value)
page.
< 0, 5, 10, 15, ................, 45> min
< 3, 3.5, 4, 4.5, ................, 19> cmH2O
< 4, 4.5, 5, 5.5, ................, 20> cmH2O
< 1, 2, 3, 4, 5, 6, 7, 8> if P set at 4-18 cmH2O
< 1, 2, 3, 4, 5> if P set at 18.5-20 cmH2O
< XXXX.X > hr
< XXXX.X > hr
< XXX >
< XXX >
10