Guidance And Manufacturer's Declaration- Electromagnetic Emissions; Guidance And Manufacturer's Declaration- Electromagnetic Immunity - Apex XT Fit Instructions D'utilisation

Masquer les pouces Voir aussi pour XT Fit:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 65
English
English

Guidance and Manufacturer's Declaration- Electromagnetic Emissions:

This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Emissions Test
Harmonic emissions
IEC61000-3-2
Voltage fluctuations /
Flicker emissions
IEC61000-3-3

Guidance and Manufacturer's Declaration- Electromagnetic Immunity:

This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Immunity Test
Electrostatic
Discharge(ESD)
IEC61000-4-2
Electrical
fast
burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltage
dips,
interruptions and voltage
variations on power supply
input lines
IEC61000-4-11
Power frequency
(50/60Hz) magnetic field
IEC61000-4-8
NOTE: U
is the a.c. mains voltage prior to the application of the test level
T
Appexdix A: EMC Information
Compliance
Class A
Complies
IEC60601 test level Compliance
±6kV contact
±8kV air
transient/
±2kV
for
supply line
±1kV for input/out
line
±1kV for differential
mode
±2kV for common
mode
short
<5 % UT(>95 % dip
in UT)for 0,5 cycle
40 % UT(60 % dip
in UT)for 5 cycles
70 % UT(30 % dip
in UT)for 25 cycles
<5 % UT(>95 % dip
in UT)for 5 sec
3 A/m
Electromagnetic Environment-Guidance
The device is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network.
±6kV contact
±8kV air
power
±2kV
for
power
supply line
±1kV for input/out
line
±1kV for differential
mode
±2kV for common
mode
<5 % UT(>95 % dip
in UT) for 0,5 cycle
40 % UT(60 % dip
in UT) for 5 cycles
70 % UT(30 % dip
in UT) for 25 cycles
<5 % UT(>95 % dip
in UT) for 5 sec
3 A/m
17
Electromagnetic
Environment-Guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least 30
%.
Mains power quality should be
that of atypical commercial or
hospital environment
Mains power quality should be
that of atypical commercial or
hospital environment.
Mains power quality should be
that of atypical commercial or
hospital environment. If the
user of this device requires
continued
operation
power mains interruptions, it is
recommended that the device
be
powered
from
uninterruptible power supply or
a battery.
Power
frequency
magnetic
fields should be at levels
characteristic
of
location in a typical commercial
or hospital environment.
during
an
atypical

Publicité

Table des Matières
loading

Produits Connexes pour Apex XT Fit

Ce manuel est également adapté pour:

9s-005500

Table des Matières