Maintenance; Daily Service; Cleaning The Cutting Chamber; Weekly Service - Struers Exotom-100 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Daily Service

Cleaning the Cutting Chamber

Weekly Service

Monthly Service

Replacing the Cooling Water

42
Exotom-100
Instruction Manual

5. Maintenance

Clean the cutting chamber, especially the cutting table and
the T-slots.
To ensure a longer lifetime for your Exotom-100, Struers strongly
recommends daily cleaning of the cutting chamber with the
flushing hose:
Lift the flushing hose from the holder on the rear wall. The
recirculation pump is activated.
Point the flushing hose towards the bottom of the cutting
chamber.
Press the lever on the flush nozzle to turn the water on.
Adjust the flow using the lever.
Clean the cutting chamber thoroughly.
Turn off the water by releasing the lever on the flush nozzle.
Do not clean the lamp glass with alcohol.
Refill the Recirculation Cooling Unit. The water level should
be 5-10 cm below the upper edge of the tank.
Clean the cutting chamber thoroughly.
Clean the filter in the filter drawer.
Check and if necessary empty and clean the sieve in the
drawer of the Recirculation Cooling Unit.
Always maintain the correct concentration of Struers Additive in the cooling
water (percentage stated on the container of the Additive). Remember to
add Struers Additive each time you refill with water.
Replace the cooling water in the Recirculation Cooling Unit at
least once a month.
Fill the tank with 150 l of water. The water level should be
5-10 cm below the upper edge of the tank.
Add Struers Additive according to the instructions.
IMPORTANT
Use a soft, damp cloth.
IMPORTANT

Publicité

Table des Matières
loading

Table des Matières