CALIBRATION

The scale can be calibrated using the following procedure. To enter this
procedure it is necessary to use Func 4 which is accessible using the
[Func/C] key as described in section 7.4 in the full manual, or by using the
passcode access as described in section 12.0.
The scales calibrate using either metric or pound weights depending on the
weighing unit being used before calibration. The display will show either "kg"
or "lb" to identify the weights expected.
PROCEDURE
• Enter the calibration section using Func 4, C8 CAL or using the
passcode as described in section 12.0.
• The display will show "unLoAd".
• Remove any weight from platform and when the stable symbol is
on press [Z/T].
• The display will show "Ld" then "0000XX" which shows the last
calibration weight used. Place this calibration weight on the scale
and press the [Z/T] key. If the weight you put on the scale does
not match the value displayed, press the [Func/C] key to clear the
vale then use the [Unit/] key and [Pcs/ ] key to set the correct
value. When it is correct press [Z/T].
• If the calibration is acceptable the scale will run a self-test during
which the calibration weight should be removed. If an error
message "FAiL L" is shown try calibration again as a disturbance
may have prevented a successful calibration.
After calibration the scale should be checked to verify the calibration is
correct. If necessary repeat the calibration making sure that the scale is
stable before accepting any weight.