RS-232 SPECIFICATION
The RS-232 is not available in WBW scales.
4.0
CALIBRATION
The WBW-M scales should not be calibrated by the user. Calibration of the scales may make it illegal to use the
scales. Contact your local metrology standards office for further assistance.
The WBW non-approved scales are calibrated using metric weights when the weighing unit selected is either
kilograms or grams and using pound masses when the weighing unit selected is either pounds, ounces or
pounds:ounces.
To start the calibration turn the scale off and then turn it on again. Press [Tare] during the self-test. Scale will show
"P- - - - "
. Enter code number "0000" using the numeric entry method (see section2.1) and press [Tare]. This will take
you directly to the calibration section.Display will show
Remove all weight from the pan and then press the [Tare] key when the scale is stable. After the Zero point is set,
"Ld xx"
the display will show
full capacity of the scale. If the mass is different from the displayed value, enter the value of the mass in whole
numbers using the numeric entry method (see section 2.1). The kg or the lb symbol will be on to show the active unit.
Press the [Tare] key when the stable indicator is on. The scale will calibrate to the mass and then return to weighing.
Remove the calibration weight as soon as calibration is complete.