Guidance and Manufacturer's Declaration- Electromagnetic Immunity:

This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Immunity Test Levels
Basic EMC
Professional healthcare
standard
facility environment
Electrostatic
Discharge
±8kV contact
(ESD)
±15kV air
IEC61000-4-2
Electrical fast
±2kV for power supply line
transient/ burst
±1kV for input/output line
IEC61000-4-4
± 1 kV line(s) to
Surge
line(s)
IEC61000-4-5
± 2 kV line(s) to earth
Voltage dips,
short
Voltage Dips:
interruptions
i) 100% reduction for 0.5 period,
and voltage
ii) 100% reduction for 1 period,
variations on
iii) 30% reduction for 25/30 period,
power supply
Voltage Interruptions:
input lines
100% reduction for 250/300 period
IEC61000-4-11
Power
frequency
(50/60Hz)
30 A/m
magnetic field
IEC61000-4-8
3 Vrms
0,15 MHz – 80 MHz
Conducted RF
6 Vrms in ISM bands
IEC 61000-4-6
between 0,15 MHz
and 80 MHz
80 % AM at 1 kHz
HOME
Compliance
HEALTH
Levels
CARE
ENVIRO
NMENT
±8kV contact
±15kV air
±2kV for
power
supply line
±1kV for
input/output
line
± 1 kV line(s) to
± 1 kV line(s)
line(s)
to
line(s)
30 A/m
30 A/m
3 Vrms
0,15 MHz – 80
MHz
6 Vrms in ISM and
amateur radio
6Vrms
bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
Electromagnetic
Environment-Guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be at
least 30 %.
Mains power quality should
be that of atypical
commercial or hospital
environment
Mains power quality should
be that of atypical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
this device requires
100-240V
continued operation during
power mains interruptions, it
is recommended that the
device be powered from an
uninterruptible power supply
or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Portable and mobile RF
communications equipment
should be used no closer to
any part of this device,
including cables, than there
commended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.