iCH Auto/English
NOTE: The minimum pressure is only to be prescribed by a doctor.
(7) Initial Pressure (only APAP mode)
Press "MENU" button to select [Init.
can view the current pressure setting displayed in cmH
adjustable only by the provider, a respiratory therapist or doctor.
NOTE: The initial pressure is only to be prescribed by a doctor.
(8) PVA (Pressure Variation Algorithm)
Press "MENU" button to select [c
enabled by the provider, a respiratory therapist or physician. The PVA function delivers
pressure relief upon exhalation to improve comfort. If the PVA function is enabled, press
"UP" or "DOWN" button to adjust the comfort level. There are 3 adjustable levels of
pressure relief that you feel while exhalation. Level 1 provides small pressure variation; and
level 3 provides higher pressure variation.
(9) Low Pressure Auto Off (only CPAP mode)
Press "MENU" button to select [LOW P on/off] menu, you can view if the Low Pressure Auto
Off setting is enabled and the device will be automatically turned off while detecting a large
leak for 3 minutes. This setting is adjustable only by the provider, a respiratory therapist or
physician.
(10) Compliance Meter
Press "MENU" button to select the [CM XXXX.X hr] menu. The compliance meter records
the total therapy hours for the device. The compliance meter should be re-set only by the
provider, a respiratory therapist or by a doctor.
(11) Alarm
Press "MENU" button to select the [Alarm on/off] menu, and press "UP" or "DOWN" button
to set the alarm on or off. When the alarm is turned on, the audible alarm will activate with
warning messages shown on the LCD display. Set alarm off to mute the audible alarm.
(12) Clock Alarm
Press "MENU" button to select the [Clock Alarm on/off] menu and press "UP" or "DOWN"
button to set the clock alarm on or off. When the clock alarm is set on, the display will show
the time on the left side. Press "UP" or "DOWN" button to set the time to wake you up. Once
the clock alarm is activated, press the start/standby button to mute the audible alarm.
(13) Clock
Press "MENU" button to select the [Clock XX:XX] menu, press "UP" or "DOWN" button to
set the current time.
NOTE: If you set a new time that occurs in the past then the "Invalid Data" which does
(14) Turning off the Device
Remove the power cord from the electrical outlet, and disconnect power cord from the power
socket on the right side of the device.
NOTE: Once the setting is confirmed, press the "MENU" button. Otherwise, the device
not exist for this new period would be erased.
will automatically go back to standby without saving the modification if no action
is taken in 20 seconds.
XX.XcmH
O] or [Init.
2
O or hPa unit. Initial pressure is
2
on/off] menu, you can view if the PVA function is
XX.XhPa] menu. Here you
11