Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 68

Specifications

Specifications are subject to change without notice.
Specification
Dimensions (W x D x H)
Enclosure
Weight
Mounting
Protection class
Pollution degree/installation
category
Power requirements
Operating temperature
Operating humidity
Storage temperature
4–20 mA outputs
Alarm relay outputs
Digital inputs
Fuses
Fittings
Table 3 General specifications
Details
452 x 360 x 804 mm (17.8 x 14.2 x 31.7 in.)
Rating: NEMA 4x/IP56 (with closed doors)
Material: PC/ABS case, PC door, PC hinges and latches, 316 SST hardware
Indoor use only. Keep away from direct sunlight.
20 kg (45 lb) without reagents and standards, 36.3 kg (80 lb) with reagents
Wall, panel or table
I
2/II
AC: 100–240 VAC, 50/60 Hz
AC instrument: 0.5 A nominal, 8.3 A maximum
Accessory output: 100–240 VAC, 5.0 A maximum
DC: 24 VDC
DC instrument: 1.2 A nominal, 9 A maximum
Accessory output: 24 VDC, 1.6 A maximum
Connection: 0.82 to 1.31 mm
stranded recommended; field wiring insulation must be rated to 65°C (149 ºF)
minimum, insulation rated for wet locations, 300 V minimum.
5 to 50 °C (41 to 122 °F)
5 to 95% non-condensing
–20 to 60 °C (–4 to 140 °F)
Four; load impedance: 600 Ω maximum
Connection: 0.33 to 1.31 mm
20 AWG) recommended, twisted pair shielded wire
Four; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC
maximum
Connection: 0.82 to 1.31 mm
stranded recommended, insulation rated for 300 V minimum and for wet
locations to maintain enclosure ratings.
Six; connection: 0.33 to 1.31 mm
20 AWG) stranded (isolated DC voltage input or an open-collector/relay contact
closure input) recommended, insulation rated for 300 V minimum and for wet
locations to maintain enclosure ratings.
Input power—AC: T 1.6 A, 250 VAC; DC: T 6.3 A, 250 VAC
Output power—AC: T 5.0 A, 250 VAC; DC: T 1.6 A, 250 VAC
Alarm relay outputs: T 5.0 A, 250 V
Sample line and sample bypass drain: 6 mm OD push-to-connect fitting for
plastic tubing
Air purge air inlet: 6 mm OD push-to-connect fitting for plastic tubing
Chemical and case drains: 11 mm (7/16 in.) ID slip-on fitting for soft plastic tubing
2
(18 to 16 AWG) wire, 0.82 mm
2
(22 to 16 AWG) wire, 0.33 to 0.52 mm
2
(18 to 16 AWG) wire, 0.82 mm
2
(22 to 16 AWG) wire, 0.33 to 0.52 mm
2
(18 AWG)
2
(22 to
2
(18 AWG)
2
(22 to
English 21

Publicité

Chapitres

Table des Matières
loading

Table des Matières