Cleaning and Disinfecting

Note: For cover care instructions, refer to the manual provided with the cover.
Warnings:
- DO NOT allow water or cleaning solution to enter the cushion. Make sure the inflation valve(s) are closed.
Sensor Ready Cushion: Make sure the Smart Check is detached from the cushion. Water in a cushion
with Sensor Ready (SR) technology can result in damage to the Smart Check and cause a false reading
when using Smart Check.
- Cleaning and disinfecting are separate processes. Cleaning must precede disinfection. Before use by a
different individual: clean, disinfect, and check the product for proper functioning.
- Clean the product regularly and when it becomes soiled.
- Follow all manufacturer bleach container and germicidal disinfectant safety guidelines.
- Note about sterilization: High temperatures accelerate aging and will damage the product assembly.
The products featured in this manual are not packaged in a sterile condition, nor are they intended
or required to be sterilized prior to use. If institution protocol demands sterilization: First, follow the
cleaning and disinfecting instructions; then, open the inflation valve and use the lowest possible
sterilizing temperature, but not to exceed 70°C (158°F), for the shortest time possible. DO NOT
autoclave.
Cautions:
- Use of the following may cause damage to the cushion: abrasives (steel wool, scouring pads); caustic,
automatic dishwasher detergents; cleaning products that contain petroleum or organic solvents, including
acetone, toluene, Methyl Ethyl Ketone (MEK), naphtha, dry cleaning fluid, adhesive remover; oil-based
lotions, lanolin, ozone gas; ultraviolet light cleaning methods.
- DO NOT expose to direct sunlight.
Hand Wash and Hand Disinfect the Cushion (Preferred Method)
Hand Wash the Cushion: Remove the cover, close the inflation valve(s), and place in
a large sink. Hand wash, using mild liquid hand soap, hand dishwashing liquid, laundry
detergent, or a multipurpose detergent (Follow product label instructions.) Gently scrub all
cushion surfaces, using a soft plastic bristle brush, a sponge or a cloth. Rinse with clean
water. Air dry.
Hand Disinfect the Cushion: Follow the hand wash instructions, using 1 part
household liquid bleach per 9 parts water. Keep the product wet with the bleach
solution for 10 minutes. Rinse thoroughly with clean water. Air dry.
Caution: Rinse thoroughly. Detergent residue may cause air cells to stick together.
7
+