Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 28

Cleaning

Toto Lateral Turning System Platform Cover
Keeping a cleaning record is recommended as the product is suitable for re-use.
Light Soiling
In the absence of gross contamination or unusual risk, the removal of dirt and spillages with a
!
neutral detergent and warm water should be sufficient. A 0.1% chlorine-based solution (1,000ppm)
may also be used. Ensure the cover is rinsed with clean water and wiped down with a single use,
non-abrasive cloth and thoroughly dried. Ensure adequate ventilation.
Heavy Soiling
In cases of gross contamination, the Toto lateral turning system platform cover should be
!
cleaned first with detergent and water and then with a 0.1% chlorine-based solution (1,000ppm).
Ensure the cover is rinsed with clean water and wiped down with a single use, non-abrasive cloth
and thoroughly dried. Ensure adequate ventilation.
If required, a 1% chlorine-based solution (10,000ppm) and cold water may be applied for no
!
more than two minutes.
Machine Washing Instructions
Disinfection may be achieved by laundering at temperatures not exceeding 65ºC for 10
!
minutes or 73ºC for three minutes
1
Protect, Rinse and Dry - BHTA guidance on the care, cleaning and inspection of healthcare
mattresses. (January 2012 - HHTA HM 0112)
Drying
To avoid shrinkage, line dry in a clean, indoor environment. Covers must be thoroughly dried
before re-fitting to the Toto platform.
Do not mangle or iron.
Platform, Cabling, Control Unit and Air Hoses
Before cleaning, disconnect from mains power supply.
The platform, cabling, control unit and air hoses can be wiped clean with alcohol wipes and/or
a chlorine derivative. Care should be taken when wiping the control unit not to allow liquids to
enter through any apertures.