Care
Inspection and Care
The Toto lateral turning system cover, platform and control unit has a unique batch number which
should be retained and used in the event of a warranty claim. NEVER remove these labels.
Cover
Regularly inspect the outer and inner surfaces for signs of damage.
Damaged covers should be reported to the ward or department manager or relevant healthcare
professional.
Torn covers should be withdrawn from service and replacements sought.
Zip fasteners should also be checked for function and integrity.
Platform
Regularly inspect the interior and exterior of the platform, ensuring the hinges and air cells are
intact.
Ensure the platform has not drifted out of position and the air hoses are not blocked, kinked or
twisted.
Fitting / Replacing Damaged Air Cells
In the event of an air cell puncture or leak, it is possible to replace either or both air cells. Please
contact Frontier Therapeutics Customer Services for replacement parts Tel: +44 (0) 1495 235800.
1. Switch off the control unit and disconnect from mains power.
2. Disconnect the air hoses to the platform using the quick connectors.
3. Unzip and pull back the cover revealing the platform.
4. Determine the faulty air cell and remove by unfastening from the Velcro®.
5. Inspect the platform for other signs of damage, including the potential source of the damage.
6. Using the Velcro®, fit the new air cell ensuring it is in-line with the correct section.
7. Replace the cover ensuring correct orientation and feed air cell hoses through apertures at the
bottom.
8. Reattach the hoses using the quick connectors - 'Red to Red' - ensuring they are not twisted or
kinked.
18