9.2 Temperature Effect
Readings from an oxygen analyzer/monitor will typically hold calibration and read
correctly within ± 2% when in thermal equilibrium with their operating environment.
Use caution to ensure the oxygen analyzer/monitor is thermally stable when calibrated
to ensure optimal performance and accurate readings by the analyzer/monitor. Maxtec
recommends the following:
• Allow adequate time for the sensor to equilibrate to a new ambient temperature.
• For best results, perform the calibration procedure at a temperature close to the
temperature where analysis will occur.
9.3 Pressure Effect
Readings from an oxygen analyzer/monitor are proportional to the partial pressure
of oxygen. The partial pressure is proportional to the absolute pressure at any given
concentration. Thus the readings are proportional to the concentration if the pressure is
held constant. Flow rate of a gas can affect pressure at the sensor due to a change in
the sensor's sensing point. Maxtec recommends the following:
• Calibrate analyzers/monitors used with the MaxFLO
sample gas.
• When a sensor is placed in the mixed gas analysis port use the same flow rates when
measuring as when calibrating.
9.4 Humidity Effect
Humidity (non-condensing) has no effect on the performance of the MaxFLO
than diluting the gas.
• Use caution to avoid condensing humidity when in operation.
10.0 SPARE PARTS AND ACCESSORIES
10.1 Included with Your Unit
Part Number:
R223P02
R219P32
R110P10-001
R223M02
10.2 Spare Parts
Part Number:
R110P10-001
R219P32
R219P33-001
R219P33-003
10.3 Optional Accessories
Part Number:
R217P62
R217P72
R213P10
R213P05
R125P03-002
R129P12
R100P49-001
R206P75
* An electronic version of this Operating Manual and Instructions for Use is available on-line at
www.maxtec.com
8
Item:
MaxFLO
, High Flow
2
22mm Male/15mm Female Conical Fitting, High Flow
Flow Diverter
Operating Manual and Instructions for Use (English)*
Flow Diverter
22mm Male/15mm Female Conical Fitting, High Flow
High Flow Valve Replacement Kit
High Flow Overhaul Kit
Item:
MaxO2 + A (Handheld Oxygen Analyzer)
MaxO2 + AE (Handheld Oxygen Analyzer)
OM-25AE (Oxygen Analyzer)
OM-25ME (Oxygen Monitor)
MAX-250E (Oxygen Sensor Medical)
2' Blender Hose
SmartStack I.V. Stand (Single, Premium)
Pole Mount (For use with Analyzers/Monitors with external sensors)
866.4.Maxtec www.maxtec.com
866.4.Maxtec
at the same pressure as the
2
www.maxtecinc.com
other
2