Abbreviation Guide; Specifications; Factors Influencing Calibration When Using The Mixed Gas Analysis Port; Flow Rate Effect - Maxtec MaxFLO2 High Flow R223P02 Manuel D'utilisation Et Mode D'emploi

Le maxflo2 est conçu pour fournir un débit intermittent ou continu contrôlé d'un mélange air/ gaz d'oxygène à un enfant, en pédiatrie et chez les patients adultes
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

should be overhauled and serviced a minimum of every four (4) years with
R219P33-003 to replace all o-rings and valves.
When using a medical grade air source, a watertrap/filter is recommended to be
attached to the air inlet of the MaxFLO
may compromise the function of the MaxFLO
Store the MaxFLO
2

7.0 ABBREVIATION GUIDE

FiO
2
°C
°F
CGA
DISS
O
2
LPM
atm
PSIG
kPaG

8.0 SPECIFICATIONS

Weight (unpackaged)
Internal Filter (air and O
O
Concentration Adjustment Range
2
Accuracy of O
Concentration
2
Gas Supply Pressure
Outlet Flow Range
Optimal Working Pressure
Maximum Allowable Pressure
Temperature Range (Mixer only)
Maximum Allowable Temperature (Mixer only)
Relative Humidity Range
Ambient Storage Conditions
Temperature Range
9.0 FACTORS INFLUENCING CALIBRATION WHEN USING THE MIXED
GAS ANALYSIS PORT

9.1 Flow Rate Effect

In order to get accurate readings from the mixed gas analysis port, for mixed gas
concentrations, the total flow must be at least 10 LPM.
in a clean, dry place when not in use.
inlet)
2
866.4.Maxtec www.maxtec.com
866.4.Maxtec
prior to use. Contaminants from gas sources
2
.
2
Fraction concentration of inspired oxygen
Degrees Celsius
Degrees Fahrenheit
Compressed Gas Association
Diameter indexed safety system
Oxygen
Liters per minute
Standard atmosphere
Pounds per square inch gauge
Kilo-pascal gauge
2.4 lbs (1.1 kg)
45 - 90
µm particulate
21% - 100% O
+/- 5% of the indicated value
25 - 75 PSIG (170 - 520 kPaG)
10 LPM (at 20.9% O
to 100 LPM (at 60% O
flow mixer
50 PSIG (345 kPaG)
100 PSIG (689 kPaG)
59°F - 104°F (15°C - 40°C)
150°F (65°C)
0 - 95% non-condensing
5°F - 122°F (-15°C - 50°C)
www.maxtecinc.com
2
and 100% O
2
) for high
2
)
2
7

Publicité

Table des Matières
loading