Component Description; Symbol Guide; Operational Requirements; Operating Instructions - Maxtec MaxFLO2 High Flow R223P02 Manuel D'utilisation Et Mode D'emploi

Le maxflo2 est conçu pour fournir un débit intermittent ou continu contrôlé d'un mélange air/ gaz d'oxygène à un enfant, en pédiatrie et chez les patients adultes
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

1.3 Component Description

Oxygen Flowmeter - An acrylic flowmeter with a stainless steel float ball with black
lettering on a white background designed specifically for the flow of oxygen.
Medical Air Flowmeter - An acrylic flowmeter with a stainless steel float ball with
black lettering on a white background designed specifically for the flow of air.
Flow Control Valve - Valve controlling the flow of gas into the flowmeter.
Mixed Gas Outlet - A diameter indexed (DISS) oxygen fitting with a 22mm
male/15mm female conical fitting.
Oxygen Gas Inlet - Fitting for connection of an oxygen inlet hose from the oxygen
gas source.
Medical Air Gas Inlet - Fitting for connection of a medical air inlet hose from the
medical air gas source.
Mixed Gas Analysis Port - Integrated port designed to divert a gas sample to an
Analyzer/Monitor for intermittent or continuous gas sampling.
IV Pole Clamp - Clamp for securing the MaxFLO

1.4 Symbol Guide

The following symbols and safety labels are found on the MaxFLO
Attention, consult accompanying
documents
Manufacturer
SN
Serial Number
REF
Catalog Number
LATEX
Latex Free
Authorized representative in
EC REP
the European Community
Federal law (USA) restricts this device
only
to sale by or on order of a physician
!
Warning

1.5 Operational Requirements

Pressurized Oxygen – For optimal performance the oxygen source must provide clean,
dry, medical grade oxygen at a line pressure of 50 ± 5 PSIG (345 ± 35 kPaG).
Pressurized Medical Air – For optimal performance the air source must provide clean,
dry, medical grade air at a line pressure of 50 ± 5 PSIG (345 ± 35 kPaG).
Note: Delivered gas characteristics are not affected by variations in input gas pressures.

2.0 OPERATING INSTRUCTIONS

2.1 Installation of MaxFLO
Note: Prior to placing the MaxFLO
Section 3 of this operating manual should be completed.
2
to an IV Pole
2
into clinical use, the performance check described in
2
866.4.Maxtec www.maxtec.com
866.4.Maxtec
to a vertical pole.
2
Read Float Ball On Center
LPM
Liter per minute flow
Gas Inlet
Gas Outlet
Maximum allowable
temperature
Maximum allowable pressure
Use no oil
Do Not
www.maxtecinc.com
.
2

Publicité

Table des Matières
loading

Table des Matières