6.2.5. Timer (C.A 6525)

Measurement range : 0 to 15 min
Calibre
0:00 to 15:00
Resolution
1 second
Accuracy
0.5 % L

6.3. Power supply

The instrument is powered by 6 x 1.5 V alkaline batteries, type LR6.
Measurement
Average*
consumption
charge life
Voltmeter
25 mA
Resistance
50 mA
Insulation
160 mA
250 V (R = 250 kΩ)
Insulation
200 mA
500 V (R = 500 kΩ)
Insulation
350 mA
1000 V (R = 1000 kΩ)
Continuity
230 mA
* Add approximately 45 mA when the backlighting is on.

6.4. Climatic conditions

Storage without battery
90
80
70
60
50
40
30
20
10
0
-50
-40
-30
-20
-10
0
10

6.5. Variations in nominal field of use

Range
Influence
for
quantities
use
Temperature
-10 to +50°C
Relative
20 to 80% RH
humidity
Supply
6.9 to 10 V
voltage
Capacity
0 to 5 μF
in parallel
at the nominal
on the resistance
current
Average
57,600 5-second measurements
28,000 5-second measurements
7,200 5-second measurements
3,800 5-second measurements
1,440 5-second measurements
3,300 5-second measurements
Use
Reference
20
30
40
50
60
70
80
Temperature in °C
Measurement
variations
Typical
Maximum
(1% R ±1 ct)/10°C(2%R ±2 ct)/ 10°C
2% R ± 2 ct
3% R ± 2 ct
(1% R ± 1 ct)/V
(2 R ± 2 ct)/V
Negligible
1% ±1 ct
38
90