6.5. Variations in nominal field of use

Influence
quantities
Temperature
Relative
humidity
Supply
voltage
Frequency
(inV and in mA)
Capacity
en parallel
on the resistance
Radiated
electromagnetic
field
(1) amplitude-modulated at 1 kHz, by 80%

6.6. Limits

The instrument is protected across all ranges against a constantly
applied voltage, between any two terminals, of 600 V AC/DC on
the C.A 6531, and 720 V AC/DC on the C.A 6533.
The C.A 6531 is protected for 10 s against an accidental
overvoltage of 720 V across all the ranges.
The current input on the C.A 6531 accepts 0.63 A, beyond which
it is protected by fuse.

6.7. Construction specifications

Overall dimensions of the unit (L x l x h): 211 x 108 x 60 mm
Dimensions of the display: 73 mm x 54.3 mm
Weight: approx. 835 g
Materials:
- Polycarbonate housing
- Crystal polycarbonate screen
- Elastomer external mouldings
- Silicon keyboard.
Stand:
Enables the instrument to be tilted at 30°. It clips onto the
bottom of the housing when not in use.

6.8. Compliance with international standards

Electrical safety: NF EN 61010-1 + A2 (Nov. 95),
Dual insulation:
Pollution level: 2
Installation category: III
Rated voltage: 600 V
Operating
range
limits
-10 to + 55°C
20 to 80% HR
7 to 10 V
DC and
15..400 Hz
0 to 5 µF at
nominal current
3 V/m from
80 to
1000 MHz (1)
NF EN 61057 (Feb. 97) and DIN EN 61557
Measurement
variations
Typical
(1% R ±1 ct)/10°C (2% R ±2 ct )/ 10°C
2% R ± 2 ct
(1% R ±1 ct)/V
1% R ±1 ct
Negligible
-
40
Maximum
3% R ± 2 ct
(2% R ±2 ct)/V
2% R ±2 ct
1% R ±1 ct
3% R
for resistance