Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Error
ERR 0a
Checksum program
ERR 0b
Checksum EEPROM
ERR 0c
Date error
ERR 0d
Communication Inter Micro-controller
ERR 1
Heating problem
I(0) > 0.5 A or < -0.5 A
ERR 2
I(0) = zero strength
Im > 150 A or < 50 A (Im = average of
ERR 3
the last 20 points of the plateau)
Im = maximum strength (plateau)
ERR 4
I > 0.8 x Im after vibrations
ERR 5
If > 0.2 x Im
If = final strength
ERR 6
Printer does not respond
ERR 8
Incoherence calibration measure strength
ERR 9
Incoherence calibration temperature at
35°C.
ERR 10
No production of iodine during phase 3
User's manual 
12/2016
SDmatic

■ Error messages

Meaning
Possible causes and solutions
Software error.
❏ Contact your Chopin agent.
Memory error.
❏ Contact your Chopin agent.
Set the date of the SDmatic.
Hardware error.
❏ Contact your Chopin agent.
Water missing / measurement probe out of order
/ resistor out of order.
❏ Verify that the reaction bowl contains water.
❏ Contact your Chopin agent.
Thiosulfate missing / measurement probe dirty.
❏ Add thiosulfate.
❏ Clean the probe.
❏ Contact your Chopin agent.
Measurement probe dirty / the stirrer does not
rotate.
❏ Clean the probe.
❏ Verify if the stirrer rotates.
❏ Verify if all reagents have been used.
❏ Verify the values measured by the probe in the
control menu.
❏ Contact your Chopin agent.
Flour not poured into the bowl / products missing.
❏ Verify that the vibrator works.
❏ Verify if all reagents have been used.
❏ Verify the values measured by the probe in the
control menu.
❏ Contact your Chopin agent.
Measurement probe dirty.
❏ Verify the values measured by the probe in the
control menu.
❏ Contact your Chopin agent.
No printing.
❏ Verify the connections and the working order of
the printer.
❏ Contact your Chopin agent.
Aberrant value.
❏ Reserved usage
Aberrant value.
❏ Reserved usage
Mistake when preparing the solution (thiosulfate
in excess).
❏ Carry out testing again.
GB
81

Publicité

Table des Matières
loading