Maintenance Of The Sdmatic; Periodic Maintenance; Problem Solving - Chopin SDmatic Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6.
MAINTENANCE OF
THE SDMATIC
Orifice to be cleaned
Fuse holder
User's manual 
12/2016
SDmatic
If the SDmatic is not used for a long time, keep the
measurement devices in distilled water.

6.1 Periodic maintenance

The SDmatic does not require particular preventive maintenance.
■ Cleaning
❏ Once a week, proceed to a cleaning of the apparatus using a
damp cloth.
❏ Wipe the surfaces of the oval hole through which the flour flows
out by blowing carefully (low pressure compressed air) or with a
brush.
■ Verification
Periodically verify the accuracy of the apparatus by performing tests
on known flours (control card type).
If the difference between the results is higher than 3%, check the
measurement chain
(page
necessary.

6.2 Problem solving

■ Replacement of the fuse
If the apparatus does not light up and after having verified that the
power supply is correct, replace the fuse of the SDmatic.
❏ Unplug the apparatus, then open the fuse holder with a small
screwdriver and replace the fuse.
❏ The fuse is located at the back of the apparatus behind a cache.
❏ Fuse characteristics: 5x20 T 1A 250V.
❏ If an error message is displayed, refer to the following table (next
page).
❏ For a deeper troubleshooting, use the control menu (see
page
82).
82) and contact your Chopin agent if
GB
79