G
English
1. General
The IMPRIMO
Cure is a light-polymerisation unit for light-curing resins, equipped with 2 different wave lengths (365 nm and 405 nm). The curing
®
chamber can be flushed with nitrogen to protect printed objects from oxygen and to avoid the creation of a dispersion layer during light curing.
2. Safety Instructions
Attention! Please read these notes carefully before connecting and operating the device. The operational safety and functioning of the device can only
be ensured if these safety instructions are observed.
1. The IMPRIMO
Cure uses ultra high-energy UV LED sources that can seriously damage health, especially eyes and skin; therefore avoid direct
®
or indirect exposure to these emissions even for short periods of time.
2. Although the IMPRIMO
Cure is equipped with an automatic UV LEDs shut-off system in case of accidental opening of the lid, the lid should
®
not be opened while the drying program is executed. If you need to open the lid, pause the program or stop it before opening the device.
3. The safety systems of the IMPRIMO
with any part of it, in particular the lid closure control system. In case of malfunctioning of the system or incorrectly turning on the LEDs when the
lid is open, immediately remove the power cable from the mains and contact the assistance service.
4. The IMPRIMO
Cure is powered at 230 volt, 80 watt main voltage. Do not tamper with any part of the device to avoid the risk of electric shock.
®
In case of malfunction of the device contact qualified assistance.
5. The IMPRIMO
®
Cure is equipped with a touch-screen interface and does not use mechanical commands or mechanical buttons. For this reason,
even when the oven does not radiate, it remains powered by the network. Remove the power supply to the oven whenever it is not used for long
periods, and especially if it remains unattended.
6. IMPRIMO
Cure uses the heat generated by LEDs to speed up the drying process of the resins. The treated objects and the oven itself can
®
become very hot.
7. IMPRIMO
Cure uses the heat generated by LEDs and nitrogen gas for the resin drying process. Do not leave the device unattended to avoid the
®
risk of fire.
8. The curing of resins can generate heat, especially if they are collected in large quantities. Never leave the IMPRIMO
the risk of fire.
9. During the drying process, the resins in the IMPRIMO
use the device in a well-ventilated area.
10. The fan placed on the bottom of the IMPRIMO
located at different points of the oven itself. Do not obstruct the pipes, their openings to the outside and the fan itself, in order not to damage the
oven itself.
11. The IMPRIMO
Cure is a device for professional use. Keep it in safe place accessible only to qualified personnel and to the rules described here.
®
Keep it away from children.
12. Keep the IMPRIMO
Cure away from heat sources, from flammable materials and liquids that may cause a safety risk. Do not wet the device and
®
clean it only with a dry cloth.
13. Connect nitrogen to the IMPRIMO
is set to 1.8 bar ex works.
14. Open the nitrogen pressure supply after switching on the IMPRIMO
after use.
3. General commissioning
Included in delivery are a power cable and a shelf for placing printed objects in the curing chamber.
Not included, but necessary, is nitrogen: it is recommended to use a gas bottle with nitrogen with an attached pressure reducer (0 - 10 bar). The
gas output should be adjusted to 2.5-4.5 bar. For the connection to the IMPRIMO
connector to your pressure reducer. The other end of the tube can be directly connected to the joint on the back of the IMPRIMO
Cure are intended as emergency aids: make sure not to tamper with the integrity of the UV oven nor
®
Cure could generate fumes of volatile substances harmful to health. Therefore place and
®
Cure has the task of controlling the temperature of the device through numerous air intakes
®
Cure with a maximum pressure of 5 bar. The working pressure on the pressure reducer of the IMPRIMO
®
Cure and loading the initial screen. Close the pressure of the nitrogen bottle
®
Cure a 6 mm tube is required. Please take care to use a proper
®
Cure unattended to avoid
®
Cure
®
Cure.
®
5