VOLTCRAFT VC-519 Mode D'emploi page 48

Multimètre à pince
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Proceed as follows to measure alternating currents (A
- Switch on the DMM with the dial switch (13) and select the
matching measuring range. Always start with the largest
measuring range.
- For measurements from 0 A to 20 A, select the range
"20A
", from 20 to 200 A the range "200A
up to 400 A the measuring range "400A
the symbol for alternating current "AC" now appear in the
display.
- The display is automatically set to zero when the current
probe is closed in alternating current measuring range.
Wait until the display shows zero before you start the
measurement.
- Push the current probe opening lever (4) and open the
current probe this way.
- Reach around the individual conductor to be measured
and close the current probe again. Position the conductor
centrally with the opening at the probe.
- The measured value is indicated on the display.
- Remove the current probe from the object to be measured after completion of the measure-
ment and switch off the device. Turn the dial switch to "OFF".
48
):
" and
". "A" and

Publicité

Table des Matières
loading

Table des Matières