The pre-heating time, with the machine cold, is about 10 minutes. By selecting the "Programs"
sub-menu you access the list of sterilisation cycles that the autoclave can execute:
CyCLE B
- Cycle 134°C Standard
- Cycle 134°C Prion
- Cycle 121°C Standard
- Cycle 134°C B Flash*
To scroll through the cycles and move the selection indicator press the (+) or (-) keys and
then confirm your choice with the
STANDArD CYCLES
CyCLE B
- 134°C: 4 min. sterilisation plus 14 min. drying
- 134°C B Flash*: 3,5 min. sterilisation plus 8 min. drying
- 121°C: 16 min. sterilisation plus 14 min. drying
CyCLE S
- 121°C: 16 min. sterilisation plus 8 min. drying
- 134°C wrapped: 4 min. sterilisation plus 8 min. drying
- 134°C B unwrapped: 4 min. sterilisation plus 6 min. drying
maximum recommended load per version 16: 0.5 Kg wrapped load – 0.2 kg porous load. For version 21: 0.7 Kg
*
wrapped load – 0.3 kg porous load.
NOTE: For each sterilisation cycle available, the manufacturer must carry out Works Tests in compliance with EN
13060 (par. 7.3.1). The cycles available for which the manufacturer has not carried out Works Tests must undergo
Installation Tests according to EN 13060 par. 7.4.
PrIoN CYCLE
HoW To USE THE MEMorY CArD
To use the Compact Flash card, comply with the instructions supplied on the CD. To
download the data, use the specific program supplied.
"LAST ADVANCED CYCLE" FUNCTIoN
The "LAST ADVANCED CyCLE" function is activated automatically for all the programs.
30' after the end of the sterilisation cycle, the heating elements switch off and a second
drying cycle automatically takes place when the temperature inside the chamber reaches
60°C.
ISTErILISATIoN TIMES INTErVENTIoN
You can increase the sterilisation times; the operation must be carried out by a specialist
technician.
"ALErT DISPLAY" FUNCTIoN
Quick flashing: from max to min every second when there are 4 minutes to cycle
completion.
46 46
- Ciclo at 134°C: 18 minutes sterilisation plus 14 minutes drying.
CyCLE S
- Cycle 121°C porous material
- Cycle 134°C wrapped
- Cycle 134°C unwrapped