CHArACTErISTICS

THE DEVICE IS INTENDED FOR STERILISATION USING WET SATURATED
STEAM.
Use the vacuum pump to obtain an optimal sterilisation process with correct interaction
between the steam and the load processed also in the cavities.
STERILISATION PROGRAMS
TyPE B
- 121°C or 134°C – cycle type B standard
- 134°C – cycle type B Prion
- 134°C B Flash – cycle type B fast
TEST PROGRAMS
TyPE B
- Vacuum test: checks the capacity to maintain
- Bowie & Dick test: checks steam penetration
- 121 Helix test
- 134 Helix test: check steam penetration in
The autoclave is available in 2 capacity variants in the sterilisation chamber:
- 16
- 21
The autoclave is available with the following data saving or data print-out devices:
- external printer (data print-out only)
- built-in printer (data print-out only)
- memory card (saves data with print-out via dedicated sw).
AUTOMATIC LAST ADVANCED CyCLE FUNCTION
When the cycle is complete, with the door closed, the autoclave goes in stand-by after
30 minutes. When the temperature of the chamber reaches 60° C the drying phase is
automatically repeated.
WATER FILLING SySTEMS
Automatic: with the help of a pump inside the autoclave.
Mains connection: possibility of connecting a demineralisation system, to increase
autonomy. Installation must be carried out by an authorised technician.
Manual: alternatively, the device is equipped with emergency manual filling on the rear.
36
the vacuum
in the porous load
hollow loads
TyPE S (optional)
- 121°C – porous materials
- 134°C – wrapped material
- 134°C – unwrapped material
TyPE S
- Vacuum test
- Bowie & Dick test