The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such environments.
Statement for the operational
environments
Performance features of the ME
system that have been determined
to be essential
to the performance
Warning
A list of all cables, transducers
and other accessories that are
relevant for the EMC compliance
Test
RF emissions CISPR11
Harmonic Emissions
Voltage Fluctuations/Flicker
Electromagnetic disturbances – Requirements and tests
Inside hospitals except for: near active HF surgical equipment and the RF
shielded room of an ME system for magnetic resonance imaging, where the
intensity of EM disturbances are high.
Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary,
this equipment and the other equipment should be observed to verify that they
are operating normally.
Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches)
to any part of the EN50 UNPLUGGED / mPack UNPLUGGED, including cables
specified by the manufacturer. Otherwise, degradation of the performance of
this equipment could result.
med 2014
None
See attachment
Compliance
Group 1
Class B
Passed
Passed
2020-06-02
43/48