Step 2: Press M to mark this test as a control solution test.
If you press
M
again, the "QC" will disappear and this test is no
longer a control solution test.
Step 3: Apply control solution.
Shake the control solution vial thoroughly before use. Squeeze
out a drop and wipe it off, then squeeze out another drop and
place it on the tip of the vial cap. Hold the meter to move the
absorbent hole of the test strip to touch the drop. Once the
confirmation window fills completely, the meter will begin
counting down. To avoid contaminating the control solution do
not directly apply control solution onto a strip.
Step 4: Read and compare the result.
After counting down to 0, the control solution test result will
appear on the display. Compare this result with the range printed
on the test strip vial and it should fall within this range. If not,
please read the instructions again and repeat the controlsolution
test.
109~165
CODE
Note:
• The control solution range printed on the test strip vial is for control
solution use only. It is not a recommended range for your blood
glucose level.
• See the Meter Maintenance section for important information
about your control solutions.
Out-of-range results:
If you continue to have test results fall outside the range printed on
the test strip vial, the meter and strips may not be working properly.
Do NOT test your blood. Contact the local customer service or place
of purchase for help.