Problem
The Philips BlueControl
suddenly switches off with a
long beep during treatment
and the LED indicator lights
up solid orange.
The Philips BlueControl
suddenly interrupts the
treatment and slowly flashes
blue.
The Philips BlueControl is
switched off with a long
beep and the LED indicator
flashes orange.
The LED indicator lights up
solid orange when I press
the on/off button.
The LED indicator lights
up solid red for 5 seconds
and the Philips BlueControl
beeps 3 times when I press
the on/off button.
The adapter becomes warm
during charging.
The LED indicator does not
flash green when the micro
USB plug is inserted into
the micro USB socket of the
Philips BlueControl.
Possible cause
The sensor is blocked.
The rechargeable battery is
empty.
The temperature of the
Philips BlueControl itself
is too high and the Philips
BlueControl switched off for
safety reasons.
The device measured a too
high skin temperature and
paused the treatment to
allow the skin to cool down.
The Philips BlueControl
was interrupted too many
times during treatment,
for instance because the
sensor sensed a too high
skin temperature or because
maximum treatment time of
40 minutes was exceeded.
The device temperature
is too high and the Philips
BlueControl switched off for
safety reasons.
The rechargeable battery is
empty.
This is normal.
The adapter is not inserted
or not inserted properly
into the wall socket.
Solution
Check the position of the Philips
BlueControl in the strap. Make sure the
strap or other clothing does not cover the
sensor. If the sensor is not covered but the
Philips BlueControl still switches off, the
skin temperature is too high. The Philips
BlueControl will switch on automatically
when the skin has cooled down.
The LED indicator flashes red. Charge
the Philips BlueControl (see chapter
'Charging').
Let the Philips BlueControl cool down for
10 minutes. Put the Philips BlueControl
back onto your skin and press the on/off
button to resume treatment.
No action required. The treatment will be
resumed automatically after the device
measures a skin temperature in the safe
range.
If this happens a few times, contact Philips.
Let the Philips BlueControl cool down for
10 minutes. Put the Philips BlueControl
back onto your skin and press the on/off
button to resume treatment.
Charge the Philips BlueControl (see
chapter 'Charging').
No action required.
Insert the adapter into the wall socket
properly.