Separation Distances

Warning:
The use of accessory or cables with the EntraFlo pump other than those specified may result in increased EMISSIONS or decreased IM-
MUNITY of the EntraFlo pump. See page iii for a list of accessories.
Warning:
The EntraFlo pump should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the
Entraflo pump should be observed to verify normal operation in the configuration in which it will be used.
Recommended Separation Distances Between Portable And Mobile RF Communications Equipment
The EntraFlo® Feeding Pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
or the user of the EntraFlo® Feeding Pump can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the EntraFlo® Feeding Pump as recommended below, according to the maximum output
power of the communications equipment.
Rated Maximum
Output Of Transmitter W
0.01
0.1
1
10
100
And The EntraFlo® Feeding Pump
Separation Distance According To Frequency Of Transmitter M
150 kHz to 80 MHz
d = 1.17√P
0.12
0.37
1.17
3.69
11.67
80 MHz to 800 MHz
d = 0.35√P
0.04
0.11
0.35
1.11
3.50
800 MHz to 2.5 GHz
d = 0.7√P
0.07
0.22
0.70
2.21
7.00