en
Technical data
Inrush current
Operating current (max.)
Relay rating (option)
Fuse
17.3.2 DC version
Operating voltage
Nominal power PointGard
2100/2200
Nominal power PointGard
2700
Cable specification
Ferrules
Operating current (max.)
Relay rating (option)
Fuse
17.4 Physical specifications
Enclosure material
Enclosure protection
Display
Size (LxWxD)
Weight
PointGard 2100 EC
PointGard 2200 CAT
PointGard 2x00 remote
17.5 Environmental parameters
For sensor specifications, see sensor data sheet of the
relevant sensor.
Temperature (storage)
Temperature (operation)
Relative Humidity (operation) 0 ... 95 % (non-condensing)
32
Max. 40 A at 230 V (AC)
50 Hz,
cold start 25°C
0.5 A
SPDT, 0.1 A - 5 A at 230 V
(AC)
, 50 Hz Min. 0.1 A -
5 A at 30 V (DC)
.
5x20 mm T 0.5 A L 250 V
¼" x 1¼" T 0.5 A L 250 V
10-30 V (DC)
6 W at 24 V (DC), 25 °C
12 W at 24 V (DC), 25 °C
2 x 1.5 mm² / 2 x AWG 16/7
1.5 mm² / AWG 16/7
8 mm / 0.3 in
insulated
2.5 A
SPDT, 0.1 A - 5 A at 230 V
(AC)
, 50 Hz Min. 0.1 A -
5 A at 30 V (DC)
.
5x20 T 2.5 A L 250 V
¼" x 1¼" T2.5 A L 250V
Fiber glass reinforced polyes-
ter
IP 66
Resolution 128 x 64 pixel,
back-lit (green or red)
255 x 280 x 120 mm
10 x 11 x 4,7 in
3.9 kg / 8.6 lbs
4.2 kg / 9.2 lbs
3.8 kg / 8.4 lbs
-20 ... +65°C / -4 ... +149°F
-20 ... +50°C / -4... +122°F
Relative Humidity (open
door)
Maximum operating altitude
Pressure
Pollution degree
17.6 Tightening torque for instrument
threads
Part
Torque Lbs. In.
Cable gland (M20
13 +/- 4.4
x 1.5)
Counter nut Cable
gland (1.5'')
Screw plug
4.4 ... 5.3
Screws to
lock/unlock the
cover
CatEx sensor
194 +/- 17
Connector screws
Power supply
4.4 ... 5.3
cable
Relay connector
4.4 ... 5.3
4 to 20mA connec-
1.9 ... 2.2
tor
17.7 Cable specifications CatEx remote
Cables must be suitable for the intended ambient
temperature. Screen must be hard-wired on both sides.
Cores
Cover
Outer diameter for cable
gland
Cable size
1)
Longitude
1) For further information regarding cable length and other possible
cable specifications, see the documentation of the corresponding Pol-
®
ytron
SE Ex sensing head or Polytron
17.8 Cable specifications PIR remote
Cables must be suitable for the intended ambient
temperature. Screen must be hard-wired on both sides.
Cores
Cover
Instructions for use
20 ... 90 % (non-condensing)
2000 m / 6561 ft
700 ... 1300 hPa
20.7 ... 38.4 inHg
PD2
Torque Nm
1.5 +/- 0.5
0.5 ... 0.6
22
2.5
22 +/- 2
0.5 ... 0.6
0.5 ... 0.6
0.22 ... 0.25
3 or 5, braided screen
≥ 80 %
6-12 mm / 0.2 In - 0.47 In
1 - 1.5 mm² (14-18 AWG)
Max. 30 m /100 feet (DD/DQ
sensor)
®
junction box.
4, braided screen on both
sides
≥ 80 %
|
Dräger PointGard 2000 Series