TCM 3000 BL
1.1 Use and operation
The TCM 3000 BL is a control unit for an electronic motor which is primarily used in operating theatres
for plastic surgery and dermabrasion - a surgical microprocessor-controlled motor system for drilling,
milling, grinding, etc.
The electronic motor is used for driving handpieces and angle pieces 1:1, dermatomes, Conform
cannula handpieces, tattooing handpieces or various saws.
1.2 Technical data
Voltage: ................................................................................ switchable:100 V, 115 V, 230 V/ 50 - 60 Hz
Power: ............................................................................................................................................. 60 VA
Device fuses (2 units): ...................................................................................................................... T 1A
Application component: ................................................................................................................ type BF
Motor rotational speed: ................................................................................................ 500 - 40´000 Upm
Max. motor torque: ......................................................................................................................... 6 Ncm
Dimensions: .......................................................................................... 115 x 120 x 180 mm (H x B x T)
Weight: ........................................................................................................................................... 1.8 kg
1.3
Ambient conditions for
operation
Rel. air humidity: ............................... max. 80%
Temperature: ................................... 10 to 40° C
Air pressure: ............................ 800 to 1060hPa
1.5 Device symbols
: Caution: refer to enclosed
documentation
1275 : EC conformity symbol
: Application component of type
BF
: Device protection class: II
IP68
: Watertight footpedal
2
1 Product description
1.4
Rel. air humidity: ............................... max. 90%
Temperature: ...................................... 0 to 60° C
Air pressure: ............................ 700 to 1060hPa
Ambient conditions for
transport and storage
: Operational earth
: Device class: APG
: Old electrical and electronic
equipment must be disposed