Troubleshooting

In instances of abnormal operation, alarm/error messages will be displayed on the
Control Panel. If a problem persists, please request assistance from the customer
support information services.
Alarm/Error Messages
Error Code
E.02
E.03
E.04
E.05
A.06
A.07
A.08
EzRay Air Wall (VEX-S300W) User Manual
A.0X: A problem occurred, and the system performs the correction
automatically. This alarm clears after the correction is completed.
E.0X: An error occurred. Turn the power off, and then turn it back on. If
the error persists, contact your Service Representative.
Check Parameter
X-ray Generator
X-ray Generator
System
User
4. Troubleshooting
Description
Error related to X-ray exposure is not
possible to exposure X-ray in the state while
"E.02", "E.03", "E.04", "E.05" where power is
maintained.
After X-ray exposure related error occurs,
when the equipment is turned off and turned
on, X-ray exposure is normally performed.
The system needs cooling time due to
continuous operation. (Duty Cycle: 1:60)
This alarm clears when the system
temperature goes down to normal.
The Remote Exposure Switch has been held
down. Release the switch.
The exposure button has been pressed and
released before the X-ray exposure finished
with the default exposure time. Press and
hold the exposure button for the duration of
the exposure time.
35