The switching hysteresis around the limit contact can be specified in % with the limit-
contact (X2SD) function.
Value range: 0.0 % to 30.0 %
The controller of a different module can be specifically turned off with the limit controller
function as soon as the limit range has been reached.
Value range (module on plug-in connection 1): none / all / module 2 / module 3
Through selecting the module selection function one returns to the controller parameter
module selection.
The page 1 function is used to display the specific parameters of the selected
controller.
4.4

Calibration

NOTE
The main menu for adjusting the Multronic can be accessed through operating the ↑-
key in the measuring mask.
Main menu
The calibration function is selected through placing the cursor onto the selected item
and operating the ENTER key.
Calibration module
selection
(not available at
Multronic OC!)
The redox measuring module that is to be calibrated can now be selected. To do so,
move the cursor onto the corresponding selected item and operate the ENTER key.
Redox calibration
This is where the settings required for the calibration are made.
The buffer solution setting is used to set the V-value of the reference buffer solution.
Value range: 0 mV to 1000 mV (for redox with ±-mV-Signal: ± 500 mV)
417101806_Anhang04_Redox.doc
The probe must be cleaned before every calibration (see chapter 5, maintenance).
The calibration of the redox-measuring electrode is equivalent to adapting the
measuring electrode to the redox measuring module.
Redox Measuring Module
- 38 -
Rev.: 06-07/10