The connection provided for the temperature sensor input is a 3-wire connection
system. If the input is to be changed to a 2-wire connection system, the red jumper
near the terminals on the measuring module must be moved.
Jumper between 1 and 2: 2-wire connection system
Jumper between 2 and 3: 3-wire connection system
NOTE
3.2

Electrode installation

The electrodes must be installed in such a way that they deviate a maximum of 80°
from vertical. This rule also applies for the calibration of the electrodes.
Horizontal (recumbent) or vertical from below (standing on its head) are not permissible
installation positions.
417101806_Anhang04_Redox.doc
The 3-wire connection system is preferable.
When using 2-wire connection systems the temperature measured is too high if
long cables are employed.
Redox Measuring Module
- 29 -