Display explained
After having connected the Clima Check to the HPM touch
the system continuously measures and outputs measurement
results on the display (real-time measurements).
The following 4 values are now shown on the HPM touch pro
display:
• Surface temperature (IR temperature)
• Air temperature
• Dew point temperature
• Relative air humidity
Conducting measurements
Temperature measurement via infrared temperature sensor
(IR temp)
Align the sensor horizontally with the surface to be measured.
The surface temperature is determined using the installed
infrared temperature sensor. The greater the distance to the
object, the greater the measuring range. Keep the measure-
ment distance to the object to a minimum. As a rule, the
distance must not exceed 100 cm. If the ascertained surface
temperature is below the dew point temperature shown on
the display of the HPM touch pro, this means there is conden-
sation on the surface of the object.
47