Using the Device during Surgery

en
Before surgery
During Surgery
After Surgery
32
8
Using the Device during Surgery
WARNING!
Function test
The function test must be performed prior to each device use.
CAUTION!
Equipment should be positioned such that power cable can be easily disconnect-
ed.
1. Tube set and instrument are filled with irrigation fluid. The suction system is
connected. The stopcocks are closed.
2. Press the suction START/STOP key (LED lights up).
The suction system is pre-evacuated.
Only then can the suction procedure be started with the instrument.
1. Press the irrigation START/STOP key (LED lights up).
2. Insert instrument into abdomen.
Start the irrigation process:
Operate the trumpet valve or open the stopcock.
3. Starting the suction process:
Operate the trumpet valve or open the stopcock.
NOTE!
The manufacturer recommends emptying the tube set using the pump when em-
ploying saccharine irrigation fluids (HF use). Remove fluid bags and instrument
first.
1. Press the irrigation START/STOP key and the suction START/STOP key (LED
goes out).
2. Use the power switch to turn device off.
3. Remove the tube sets.
NOTE!
Comply with hygiene rules and regulations when disposing of the tube set, col-
lected fluid, and the suction container.