Operating the Device
en
RFID technology (transponder technology)
Invalidating a tube set
Transponder signal loss
18
5.4

Using the Tube Sets

Transponder technology detects the tube type, the validity and reliability of a
tube set automatically. This eliminates virtually all "operating errors" since non
fitting, invalid, and not approved tube sets are reliably detected. In any approved
tube set (see chapter 12 "Accessory list"), the corresponding transponder is locat-
ed under the tube holder.
If an approved tube set is located in the device, the transponder technology au-
tomatically invalidates this tube set when the device is started. The tube status
LED lights up green. Depending on the tube set, this has the following conse-
quences:
• Single-use tube set: After inserting the tube set and starting the irrigation cy-
cle, the tube set is invalidated after 10 minutes. If the irrigation cycle is
stopped, the device can be restarted within 30 minutes. If the device is
switched off or in case of a power failure, the tube set is invalidated. The irriga-
tion cycle can no longer be started. The tube status LED lights up red. If this is
the case, you must insert a new, valid, and approved tube set.
• Reusable tube set (NOT FOR SALE IN USA): The reusable tube set can be used
about 20 times. After inserting the tube set and starting the irrigation cycle,
the tube set is invalidated after 10 minutes. If irrigation is stopped during a use
cycle, it can be restarted within 30 minutes without the set being invalidated.
If the device is switched off or in case of a power failure, the current use cycle
is invalidated. If an invalid reusable tube set has been inserted, a warning sig-
nal (beep) is emitted. A red LED will be displayed for 8 seconds.
• Day patient tube set also day tube set (NOT FOR SALE IN USA AND CHINA): The
following description refers only to the day tube (for the patient tube, please
see chapter 5.7 Inserting a Day Patient Tube Set (NOT FOR SALE IN USA AND
CHINA)).
The reusable day tube set can be used for a total of 10 times. After inserting the
tube set and starting the irrigation cycle, the tube set is invalidated after 10
minutes. If irrigation is stopped, it can be restarted within 30 minutes without
the set being invalidated.
After the device is switched off or in the case of a power failure, the current use
cycle is invalidated. Irrigation can no longer be started following the last use
cycle. The tube status LED lights up red. If this is the case, you must insert a
new, valid, and approved day tube.
The device stops and cannot be restarted if the transponder loses its signal dur-
ing current use. If the signal is restored within 20 seconds, it is possible to contin-
ue using the tube set.