Functional Test
en
Fig. 7-1
Irrigation test
30
7
Functional Test
WARNING!
Function test
The function test must be performed prior to each device use.
WARNING!
Preventing infections
Sterilize reusable instruments before surgery to prevent infections. Check all the
single-use/disposable items before removing them from the package to ensure
that the packaging is intact and that the expiration date is still valid.
WARNING!
Original accessories
For your own safety and that of your patient, use only original accessories.
7.1
Preparing the Function Test
1. The device is switched off and without an inserted tube set.
2. Switch device on using the power switch.
3. The device carries out a self-check. A long beep is emitted after the successful
device check.
4. Insert tube set (see Fig. 5-4 "Inserting the tube set"). Connect the suction sys-
tem (see Fig. 5-7 "Inserting the suction/irrigation set").
7.2
Irrigation Test