Table 4: Guidance and Manufacturer's Declaration - Electromagnetic Immunity
E
The
ConMed Linvatec
electromagnetic environment specified below. The customer or the user of this
VP47XXF
Immunity Test
Conducted RF
IEC 61000-4-6
VP47XX
and
VP47XXF
Monitor should assure that it is used in such an environment.
IEC 60601
Compliance Level
Test Level
3 Vrms
3 Vrms
150 kHz to 80 MHz
150 kHz to 80 MHz
Monitors are intended for use in the
Electromagnetic environment
Portable and mobile RF
communications equipment should
be no closer to any part of the
VP47XX
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended Separation
Distance
d
= [3.5 / V1]
E-44
VP47XX /
guidance
,
and
VP47XXF
Monitors
P