ConMed Linvatec VP4726 Manuel D'instruction page 40

Moniteur medical lcd 66 cm et 81 cm
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

E
Classification of Equipment
According to protection against electric
shock.
According to degree of protection against
electric shock.
According to degree of protection against
harmful ingress of water.
According to the degree of safety in the
presence of Flammable Anesthetics mixture
with air, with oxygen, or with nitrous oxide.
According to the mode of operation.
UL Safety Compliance:
EEC Safety Compliance:
NOTE: This monitor complies to the above standards only when used with the supplied
medical grade power supply, 24 Volt: Bridgepower Corporation.
The monitor should be powered from a center tapped circuit when used in the US at voltages
over 120 volts. Monitor is intended for continuous operation.
This display is energized from an external electrical power source for class 1 equipment. It is
the responsibility of the installer to test the display's earth ground to verify that it complies
with the hospital, local and national impedance requirements.
Table 2: Safety, Classifications
Class I [Grounded]. This includes the AC/DC
Adapter (Bridgepower Corporation:
JMW1150KA2400F09)
Monitor, JMW1180KA2400F01 (VP4732PS) for the
32" Monitor)
Type B
IPX1
Equipment is NOT suitable for use in the presence of
flammable anesthetics mixture with oxygen or with
nitrous oxide.
Continuous
This LCD monitor is U.L. Classified
with respect to electric shock, fire and
mechanical hazards only in accordance
with 60601-1/CAN/CSA C22.2 NO. 601.1.
This display meets the requirements of
EN-60601-1 so as to conform to the
Medical Device Directive 93/42/EEC
(general safety information).
E-36
Parameter Value
(VP4726PS) for the 26"

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Linvatec vp4726fLinvatec vp4732Linvatec vp4732f

Table des Matières