Dispenser DP 30
8
Cleaning and disinfection
The following point is important with regard to caring for the material:
● Perform cleaning and disinfection after every treatment.
8.1
Control unit and pedal
Wipe the outside using tested surface disinfectant or 70 percent of isopropyl alcohol. The front plate of the
control unit is sealed accordingly for this purpose and can be wiped clean.
8.2
Tube set, REF 6022
● Single-use tubing sets may not be reused.
● Tube sets must be disposed of properly after use.
● Don't use tube set when pack was open or damaged .
● Don't use tube set when expiry date is timed out.
● Use only Nouvag tubing sets with REF 6022!
Sterility cannot be guaranteed by reusing and re-sterilization of tubing sets. The characteristics of the
material change in a manner that can result in failure of the system. This may result in serious infections
or even patient death worst-case.
Dispenser DP 30, operating instructions, REF 31678, V52/17
EN
11