PC 1.5 Cable
This cable is required to connect the 7-pin connector to a PC (25-pin
connector).
PC 2.1 Cable
This cable is required to connect the 9-pin connector to a PC.
PC 1.2 Adaptor
This adaptor is required to connect the 9-pin connector to an 8-way
serial interface (25-pin plug).
AK 2.1 Cable
This cable is required to connect the 7-pin connector to a chart recor-
der (4mm pin plug).
AK 2.2 Cable
This cable is required to connect the 15-pin connector to a chart
recorder (4mm pin plug).
Motor protection / safety device
If the motor is locked or if a load is placed on it resulting in a higher
temperature than is permitted, the device is turned off automatically
by the safety circuit and the control light flashes. To eliminate the
error, the weight of the material must be reduced. Allow the device
to cool off.
The device must be turned off and backon again.
The motor is turned off immediately in the event of a malfunction by
a safety circuit. Amalfunction occurs if the safe and secure functiona-
lity of the device cannot be guaranteed. A malfunction is displayed
for the Basic model by the red Off signal light coming on and flas-
hing, and for the Control model by the error message being shown
on the display of the speed indicator (see the section "Error Codes".
Maintenance and cleaning
The HS/KS260basic and HS/KS260control are maintenace-free. It is
subject only to the natural wear and tear of components and their
statistical failure rate.
When ordering spare parts, please give the manufacturing
number shown on the type plate, the machine type and the
name of the spare part.
Use
only
cleansing
agents
ved by IKA to clean IKA devices. To remove use:
Dyes
isopropyl alcohol
Construction materials
water containing tenside / iso
propyl alcohol
Cosmetics
water containing tenside / iso
propyl alcohol
Foodstuffs
water containing tenside
Fuels
water containing tenside
For materials which are not listed, please request information from
IKA. Wear the proper protective gloves during cleaning of the devices.
Electrical devices may not be placed in the cleansing agent for the
purpose of cleaning.
Before using another than the recommended method for cleaning or
decontamination, the user must ascertain with the manufacturer that
this method does not destroy the instrument.
Repairs:
Please only send devices in for repair that have been cleaned
and are free of materials which might present health hazards.
For this, use the "certificate of compliance" form which you can
obtain from IKA or can download a version for printing from the IKA
website at www.ika.com.
If your appliance requires repair, return it in its original packaging.
Storage packaging is not sufficient when sending the device - also use
appropriate transport packaging.
which
have
been
appro-
33