1.- Presentation

CAMRegis is a solution for capturing, storing and recording temperatures and other physical variables such as humidity and
pressure.
They have an internal memory that can store up to 6 years' data, recording every 30 minutes, and a battery that provides up to
6 hours' autonomy recording data + up to 3 months' maintaining of date and time in the event of a failure of the electrical
supply.
The data stored can be displayed on screen or printed by means of the unit's own printer (depending on the model).
These units can be connected to each other creating networks of up to 7 loggers (1 master + 6 slaves) or connected to a PC
(master), creating a network of up to 127 units (slaves).
From the master (PC or logger) the data of any slave can be displayed and printed.

1.1.- Versions and references

MODELS
INPUTS
AKO-15740
10
AKO-15742
10
5
AKO-15750
AKO-15752
5
2
AKO-15780
2
AKO-15782
Software included: SOFTRegis

1.2.- Maintenance

Clean the surface of the alarm using a soft cloth, soap and water. Do not use abrasive detergents, petrol, alcohol or
solvents.
Units that incorporate rechargeable electrical accumulators:
This unit has built-in accumulators which must be replaced when the autonomy of the unit is less than the duration
shown in the specifications. At the end of the unit's life, the accumulators must be taken to a selective disposal
centre or returned to the equipment manufacturer.
Periodic checks:
In accordance with standard UNE EN 12830, maintenance must include the checks indicated in the UNE EN
13486 standard (only with the NTC probes supplied with the unit).

1.3.- Precautions

Using the logger contrary to the manufacturer's instructions may affect the device's safety requirements. To ensure
that the device operates correctly only probes supplied by AKO should be used.
Between –40 ºC and +20 ºC, if the NTC probe is extended to up to 1,000 m with minimum 0.5 mm2 wire, the
maximum deviation will be 0.25 ºC (wire for probe extension ref. AKO-15586).
The graph paper is thermal and therefore if you wish to keep the graphs for a long time you must make photocopies.
The length of an input's graph is less than the length of a sheet of DIN A4.
With the recording frequency or interval set to 15 minutes, the loggers keep the information in memory for over
one year. They therefore ensure compliance with UNE EN 12830, enabling graphs to be printed or displayed when
required.
If frequencies of less than 15 minutes are configured, in order to comply with UNE EN 12830, the graphs must be
printed before the memory ends and be kept for one year.
RELAYS
PRINTER
(250V, cos j=1)
NO
YES
NO
Max. alarm: 8A SPDT
Max. alarm: 8A SPDT
YES
NO
YES
22
POWER SUPPLY
100 - 240 Vac