Using the refrigerator
7.5
Alarms
A
NOTICE!
• If you are not able to eliminate the problem triggering the alarm,
store the medicinal products somewhere else as soon as possible.
• In case of power failure, the inner temperature will raise in 2 hours
(hold over time) from 2 °C to 10 °C at an ambient temperature
of 25 °C.
There are five special cases when alarms occur:
Alarm event
Temperature out of range
Power failure
Inner fan is blocked
Door open for > 20 sec
Connection problem between
display and controller
Door open for > 60 sec
The alarm can be muted and unmuted:
➤ Press the
button.
I
NOTE
• If another alarm is triggered the alarm signal is automatically
unmuted.
• After all alarm events have passed, a muted alarm signal must be
manually reset to unmuted.
22
Number of beeps
1 beep / 5 sec
1 beep / 3 sec
n.a.
1 beep / 2 sec
2 beeps / 2 sec
HC302D, HC502D
Duration of the beeps
1 sec
0.5 sec
permanent beeping
1 sec
0,5 sec
EN